Thursday, October 11, 2012

1210.2765 (J. Brack et al.)

Characterization of the Hamamatsu R11780 12 inch Photomultiplier Tube    [PDF]

J. Brack, B. Delgado, J. Dhooghe, J. Felde, B. Gookin, S. Grullon, J. R. Klein, R. Knapik, A. LaTorre, S. Seibert, K. Shapiro, R. Svoboda, L. Ware, R. Van Berg
Future large water Cherenkov and scintillator detectors have been proposed for measurements of long baseline neutrino oscillations, proton decay, supernova and solar neutrinos. To ensure cost-effectiveness and optimize scientific reach, one of the critical requirements for such detectors are large-area, high performance photomultiplier tubes (PMTs). One candidate for such a device is the Hamamatsu R11780, a 12" PMT that is available in both standard and high quantum efficiency versions. Measurements of the single photoelectron response characteristics, relative efficiencies of the standard and high quantum efficiency versions, a preliminary measurement of the absolute quantum efficiency of the standard quantum efficiency version, and a two-dimensional scan of the relative efficiency across the photocathode surface are presented in this paper. All single photoelectron investigations were made using a Cherenkov light source at room temperature at a gain of 1*10^7. These results show that the R11780 PMT is a excellent candidate for such large optical detectors, including the peak-to-valley ratios greater than 2, transit time spreads around 1.3 ns, and late-pulsing probabilities less than 5%.
View original: http://arxiv.org/abs/1210.2765

No comments:

Post a Comment